Specifications
AV DISCOVERY SystemSingle Inlet System
|
AERO VAC Gas AnalyzerMagnetic Sector Configuration2" radius, 60° deflection Resolution50-75 (with 0.010" slits) (m/D m @ 10% peak height) Source Sensitivity1x10-4 amps/Torr (1mA emission) (1x10-3 amps/Torr with 3mA emission) Minimum Detectable Partial Pressure1x10-10 Torr for Faraday Cup 5x10-14 Torr for Electron-multiplier (with 0.010" slits) Mass Range1-150, 2-250, 3-350 AMU - Standard Dynamic Range8 decades total 6 decades concurrently Emission Current0.1 - 5.0 mA adjustable |
Electron Energy25 - 120 Volts adjustable, standard Filament Power20 watts, includes Softstart feature Materials in VacuumStainless Steel (304), nickel, glass (EM), ceramics, ThO2/Ir , copper gasket ENCOUNTER Software:Operational ModesAnalog display, Histogram Display Trend Display, Leak Detect Control Panels, Calculators and Tools (See Encounter brochure for further details) Other OptionsCalibrated Standards Multiple Inlet System Wet or Dry Pump Versions RequirementsPentium III Processor, Windows 9x, 2000, XP, One Serial or Ethernet Port
Air supply for valve actuation - 80psi |