Specifications
Leak DetectorDirect/Counter-flow Mode
|
Optional RGA InformationMagnetic Sector Configuration2" radius, 60° deflection Resolution50-75 (with 0.010" slits) (m/D m @ 10% peak height) Source Sensitivity1x10-4 amps/Torr (1mA emission) (1x10-3 amps/Torr with 3mA emission) Minimum Detectable Partial Pressure1x10-10 Torr for Faraday Cup 5x10-14 Torr for Electron-multiplier (with 0.010" slits) Mass Range1-150 AMU - Low mass magnet 2-250 AMU - Standard magnet 3-350 AMU - High mass magnet Dynamic Range8 decades total 6 decades concurrently Emission Current0.1 - 5.0 ma adjustable
|
Electron Energy25 - 120 Volts adjustable, standard Filament Power20 watts, includes Softstart feature Materials in VacuumStainless Steel (304), tungsten, nickel, glass (EM), ceramics, ThO2/Ir, copper gasket, W (optional) Optional Software InformationOperational ModesAnalog display Histogram Display Trend Display Leak Detect Control Panels Calculators and Tools (See Encounter brochure for further details) Other OptionsBar Code Reader Calibrated Leaks Sniffer Probe High Pressure RGA sampling
|