Specifications
AV WATCHMAN SystemOperating Pressure RangeDirect Valve OpenBelow 10-4 Torr By-Pass Valve100 Torr to 10-4 Torr Valve ActuationDirect ValveManual, Electrical or Pneumatic By-Pass / Leak ValveManual or Automatic to preset position ProtectionPressure protection by Cold Cathode Gauge controller provides RGA filament interlock and closes direct valve if system pressure is too high (on optional automatically actuated systems). DimensionsConfiguration dependant
See next column for details on Odyssey Mass Spectrometer sensor |
AERO VAC Gas AnalyzerMagnetic Sector Configuration2" radius, 60° deflection Resolution50-75 (with 0.010" slits) (m/D m @ 10% peak height) Source Sensitivity1x10-4 amps/Torr (1mA emission) (1x10-3 amps/Torr with 3mA emission) Minimum Detectable Partial Pressure1x10-10 Torr for Faraday Cup 5x10-14 Torr for Electron-multiplier (with 0.010" slits) Mass Range1-150, 2-250, 3-350 AMU - Standard Dynamic Range8 decades total 6 decades concurrently Emission Current0.1 - 5.0 mA adjustable |
Electron Energy25 - 120 Volts adjustable, standard Filament Power20 watts, includes Softstart feature Materials in VacuumStainless Steel (304), nickel, glass (EM), ceramics, ThO2/Ir , copper gasket ENCOUNTER Software:Operational ModesAnalog display, Histogram Display Trend Display, Leak Detect Control Panels, Calculators and Tools (See Encounter brochure for further details) Other OptionsCalibrated Standards Multiple Inlet System Wet or Dry Pump Versions RequirementsPentium III Processor, Windows 9x, 2000, XP, One Serial or Ethernet Port
Air supply for valve actuation - 80psi |